Recall Z-1814-2017 ยท GWG Endoscope, Neurological

Aesculap Implant Systems LLC

AESCULAP MINOP InVent 30 Trocar System, non-sterile Product Usage:The Minop InVent Trocar System intended use is for endoscopic procedures within the central nervous system, especially for the treatment of intra- and paraventricular pathological structures.

DateMar 20, 2017 (initiated Mar 7, 2017)
StatusTerminated
ReasonAesculap Implant Systems LLC is recalling the Minop Trocar due to the possibility it may have sharp edges on the distal end which may lead to the abrasion of the insulation when removing the electrode.
Root causeNonconforming Material/Component
ActionOn March 15, 2017, 16 facilities and 1 Sales Rep were sent an Urgent Medical Device Recall Notification letter. Letters were sent Fed-Ex overnight. Customers were asked to immediately discontinue use and quarantine the product. A Sales Representative will remove the affected product and return to Aesculap Inc.
Quantity21 units distributed in U.S.
DistributionProduct was distributed throughout the United States and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983365AESCULAP MINOP SYSTEM
Aesculap, Inc., Dec 18, 1998