ClearanceBrief
Recall Z-1683-2017 ยท FTC Light, Ultraviolet, Dermatological

National Biological Corp

Hand Foot II, UVB-138. One device per box; device is surrounded by foam inserts inside a corrugated box. For UV phototherapy treatment.

DateMar 20, 2017 (initiated Jan 27, 2017)
StatusTerminated
ReasonThe wiring may allow the lamps to turn on with the key switch rather than the timer.
Root causeEmployee error
ActionIndividual patients were contacted via telephone. Replacement devices have been distributed. The distributor was contacted via email. The firm has requested that consignees return the recalled device to their firm to be held under quarantine.
Quantity7 units
DistributionWorldwide distribution. US states: IA, IL, MI, NY, PA, and WI. Country: Jordan. There are no government accounts for this recall. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K872604HAND/FOOT 624-208
National Biological Corp., Jul 24, 1987