Recall Z-1768-2017 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Nihon Kohden America Inc

Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).

DateMar 20, 2017 (initiated Mar 13, 2017)
StatusTerminated
ReasonThe Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
Root causeNonconforming Material/Component
ActionA technical bulletin was issued on March 7, 2017 to make Life Scope G9 users aware of following interaction: the CNS 6201A and CNS 9701A (using the noted software versions) will not Auto Resume when a patient is reconnected to a Life Scope G9 if the patient tile is Paused. Therefore, each time a patient is reconnected to the Life Scope G9, the Pause must be cancelled and monitoring manually resumed on the CNS 6201A or CNS 9701A. A customer letter dated March 13, 2017 was sent to all customers to inform them that Nihon Kohden America (NKA) is voluntarily notifying them that the Pause functio
Quantity654 units total (230 units in US)
DistributionWorldwide - US Nationwide Distribution including the countries of Germany, Spain, Brazil, Paraguay, Bolivia, Chile, Colombia, Costa Rica, Mexico, Peru, Pakistan, Israel, Qatar, Saudi Arabia, South Africa, Tunisia, United Arab Emirates, Korea, Malaysia, Singapore, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151080Nihon Kohden CSM 1901 Bedside Monitor
Nihon Kohden, Nov 5, 2015