ClearanceBrief
Recall Z-1825-2017 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Cincinnati Sub-Zero Products Inc

Hemotherm CE, model 400CE, Heater-Cooling Device

DateMar 9, 2017
StatusTerminated
ReasonImproperly rated fuses were used in the manufacture of the device.
Root causeComponent design/selection
ActionLetters were sent to customers and distributors instructing them to immediately cease use and/or further distribution of affected product, remove and discard the affected fuses and replace with the new fuses provided by CSZ, and complete and return the Field Action Response Form.
Quantity390 devices
DistributionWorldwide distribution. The recalled product was distributed to the following states: AL, AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The recalled product was distributed to the following countries: Mexico, Taiwan. There are no Canada accoun
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122813HEMOTHERM MODEL 400CE COOLER/HEATER SYSTEM
Cincinnati Sub-Zero Products, Inc., Dec 18, 2012