ClearanceBrief
Recall Z-1752-2017 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircula

DateApr 6, 2017 (initiated Feb 24, 2017)
StatusTerminated
ReasonThe brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Root causeNonconforming Material/Component
ActionPhilips Healthcare sent an Urgent Field Safety Notice letter informing consignees of the issue through: Field Safety Notice (FSN) CLE17-006_88200502_88200503. Customers will be notified by direct communication from Philips Field Service Engineers (FSE). A copy of the FSN will be provided to customers by the FSE during the onsite visit for the inspection. A Business Reply form is attached to the FSN for customers to acknowledge receipt and understanding of the FSN. Customers with questions were instructed to contact the Customer Care Solutions Center at 1-800-722-9377. For questions regardin
Quantity0
DistributionWorldwide distribution - US including AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, VA, WA, WI, and WV. Internationally to Australia, Austria, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, France, Germany, India, Israel, Italy, Japan, M
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062298APOLLO GAMMA CAMERA SYSTEM
Adac Laboratories, Aug 23, 2006
K080927BRIGHTVIEW VCT IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Apr 11, 2008