Recall Z-1490-2017 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
ONCOR" Avant-garde
| Date | Feb 14, 2017 |
| Status | Terminated |
| Reason | Software update |
| Root cause | Software design |
| Action | Siemens mailed an Customer Information Letter on 02/15/2017 to affected customers to inform them of a software update they plan to implement to reduce risk of collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table, using the SIMTEC-Auto Field Sequence Option. |
| Quantity | 80 systems |
| Distribution | Distributed throughout the United States |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K031764 | ONCOR AVANT-GARDE WITH COHERENCE WORKSPACES Siemens Medical Solutions USA, Inc., Sep 5, 2003 |