Recall Z-1539-2017 ยท HSB Rod, Fixation, Intramedullary And Accessories

Integra LifeSciences Corp.

Panta Nail, Rx only, Sterile,

DateMar 3, 2017
StatusTerminated
ReasonVoluntary recall/Removal of Panta and Panta XI Nails because there may be the potential that the internal and external package seal does not provide the level of sterility intended.
Root causePackaging change control
ActionIntegra sent "Urgent Medical Device Recall Letters and Return Response Forms" dated March 3, 2017 to their distributors and customers. All customers were asked to examine their inventory to determine if they had the recalled lots and if they did they are instructed to stop using them immediately. Distributors are asked if the recalled product was further distributed, those customers should be notified and instructed to follow the recalling procedures. The return response form should be completed, even if they do not have any recalled lots, annotating the amount of product and list a record of
Quantity4,449 units
DistributionUS Nationwide and Internationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050882NEWDEAL ANKLE NAIL
New Deal, S.A., May 27, 2005
K091788PANTA NAIL (DIA. 10MM), PANTA NAIL XL
Newdeal Sas, Jul 29, 2009