Recall Z-1455-2017 ยท GEX Powered Laser Surgical Instrument

Strata Skin Sciences, Inc.

XTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment of the skin conditions including psoriasis, vitiligo, atopic dermatitis, and leukoderma.

DateMar 27, 2017 (initiated Jan 23, 2017)
StatusTerminated
ReasonWhen the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.
Root causeRadiation Control for Health and Safety Act
ActionSTRATA Skin Sciences, Inc., planned action to bring defect into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 1000 domestic units and 612 international units will consist of the following: 1. Send Field Correction notification to all consignees by January 30, 2017, 2. Verify all consignees received the notification by February 10, 2017, 3. Ship all Rear Panel Boards to consignees by May 1, 2017, and 4. All corrective actions will be performed at no cost to the purchasers For further questions call Customer service phone no.: (877)
QuantityUS - 1000 , Foreign - 612
DistributionUSA (nationwide) Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.