Recall Z-1695-2017 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proteus 235

DateMar 23, 2017 (initiated Mar 9, 2017)
StatusTerminated
ReasonThe Proton Therapy System can allow gantry rotation even if the snout is not properly locked.
Root causeComponent design/selection
ActionA Field Safety Notice was mailed or handed to affected customers to inform them of the issue. The notice provides information on the issue, the risk for the patient and user and actions to be taken before starting gantry rotation. Corrective action developed by IBA are detailed.
Quantity8 worldwide, 4 in U.S.
DistributionDistributed in MA, FL, VA, and TN. China, South Korea, France and Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101508MODIFICATION TO IBA PROTON THERAPY SYSTEM-PROTEUS 235
Ion Beam Applications S.A., Aug 6, 2010