Recall Z-2577-2017 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Zimmer Biomet, Inc.
Por fullct fem st 15x200mm, Sterile, prosthesis, hip, semi-constrained, metal/polymer, porous uncemented- Specific Knee, Hip and Nail Implants
| Date | Mar 20, 2017 (initiated Feb 28, 2017) |
| Status | Terminated |
| Reason | Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures. |
| Root cause | Packaging |
| Action | Zimmer Biomet initiated a field action for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures. Zimmer Biomet sent letters via Fed-Ex and e-mail dated 02/28/2017 to customers informing them of the recall, and instructing them to do the following: 1. Review this notification and ensure affected team members are aware of the contents. 2. Complete the Certification of Acknowledgement portion of Attachment 1 a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 3. Assist the Z |
| Quantity | 9867 |
| Distribution | Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singap |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K953337 | BETA HIP PROSTHESIS Zimmer, Inc., Jan 22, 1996 |