ClearanceBrief
Recall Z-1709-2017 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Access 2 Immunoassay Systems, Part No. 81600N Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

DateMar 15, 2017 (initiated Mar 7, 2017)
StatusTerminated
ReasonBeckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.
Root causeSoftware Design Change
ActionAn Urgent Medical Device Recall letter was sent on 3/7/17 to inform customers that Beckman Coulter has determined through internal investigation that a circuit board that controls motor functions in the instrument systems listed above can affect device positioning on the pipettor and incubator belt while the instrument is running. Customers are informed of the impact, action to be taken, and the resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Customer Technical Support Center: From
Quantity851 units total for all products (291 units in US)
DistributionAffected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K922823ACCESS THEOPHYLLINE IMMUNOASSAY REAGENTS
Bio-Rad Laboratories, Inc., Aug 5, 1992