Recall Z-1760-2017 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ConMed Corporation

HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.

DateFeb 28, 2017
StatusTerminated
ReasonFor over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
Root causeMixed-up of materials/components
ActionConMed sent an URGENT: MEDICAL DEVICE RECALL Notices (dated 2/28/2017)to all affected US customers by USPS First Class Mail. International letters are in the process of translation and are scheduled to be sent by March 15, 2017. Customers were instructed to review their inventory for any of the affected product and were asked to complete the business reply form and return it to ConMed Corporation, 525 French Road, Utica, New York 13502. Customers with questions or requests should contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email hyfr2000@conmed.com. For
DistributionWorldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963088SHEATH, ELECTROSURGICAL PENCIL
Aspen Laboratories, Inc., Sep 13, 1996