Recall Z-2057-2017 ยท LHQ System, Telethermographic (Adjunctive Use)

Med-Hot Thermal Imaging, Inc.

MEDHOT MTI 2000 THERMAL IMAGING SYSTEM - TotalVision software with Elevated Body Temperature Screening module, for use as a medical diagnostic purposes as part of a telethermographic system utilizing FLIR cameras, Models A615sc and A315sc. Cameras are private labeled as MAX Basic, MAX 76 and MAX 307.

DateJan 14, 2017
StatusTerminated
ReasonIndications for use are not cleared or approved for market in the U.S.
Root causeLabeling Change Control
ActionCorrect software to eliminate any reference to temperature. Test function and notify users of software update with new user instructions.
Quantity12
DistributionUS: AL and NJ - Foreign: Nigeria
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063047MEDHOT MTI 2000 THERMAL IMAGING SYSTEM
Medhot Thermal Imaging, Inc., Oct 19, 2006