Recall Z-1467-2017 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis Zee and Zeego fluoroscopic x-ray system

DateFeb 27, 2017
StatusTerminated
ReasonA gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch.
Root causeProcess control
ActionA Customer Safety Advisory Notice (CSAN) was mailed to affected customers on 02/27/2017 to inform them of the issue. The notice provided information on what actions to take should the foot switch fail to function, the risks involved and what corrective actions they are taking to fix the issue.
Quantity2158 units distributed in US and 6,139 worldwide
DistributionDistributed Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008