ClearanceBrief
Recall Z-1743-2017 ยท NAD Dressing, Wound, Occlusive

Covidien LLC

Covidien Kerlix Super Sponge Saline Dressing, Sterile Item Code: 3338 Curity sodium chloride dressing is saturated in a 20% hypertonic saline solution. The dressing is designed to help promote biological cleansing within the wound and aid in the autolytic debridement process.

DateMar 27, 2017 (initiated Mar 3, 2017)
StatusTerminated
ReasonProduct sterility is compromised due to breach of sterile barrier
Root causePackaging process control
ActionMedtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000.
Quantity38794
DistributionWorldwide Distribution. US (Nationwide), Internationally to Canada, Australia, Brunei, Japan, Mexico, Philippines, Singapore, Andorra, Belgium, Botswana, Czech Republic, Denmark, Egypt, France, Greece, Hungary, Ireland, Israel, Italy, Kenya, Libya, Luxembourg, Mauritius, Namibia, Netherlands, Norwa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.