Covidien LLC
Covidien Curity Eye Pad Item Code: 91650 The sterile eye pad is a device that consists of a pad made of various materials, such as gauze and cotton, intended for use as a bandage over the eye for protection or absorption of secretions. It is supplied sterile to allow for use in the operating room, and for application to the eye following injury and or surgery in order to reduce the likelihood of
| Date | Mar 27, 2017 (initiated Mar 3, 2017) |
| Status | Terminated |
| Reason | Product sterility is compromised due to breach of sterile barrier |
| Root cause | Packaging process control |
| Action | Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000. |
| Quantity | 1659500 |
| Distribution | Worldwide Distribution. US (Nationwide), Internationally to Canada, Australia, Brunei, Japan, Mexico, Philippines, Singapore, Andorra, Belgium, Botswana, Czech Republic, Denmark, Egypt, France, Greece, Hungary, Ireland, Israel, Italy, Kenya, Libya, Luxembourg, Mauritius, Namibia, Netherlands, Norwa |