Recall Z-1740-2017 ยท HMP Pad, Eye
Covidien LLC
Covidien Curity Eye Pad Oval, 1-5/8 x 2-5/8 (4.2 cm x 6.7 cm), Sterile Item Code: 2841 The sterile eye pad is a device that consists of a pad made of various materials, such as gauze and cotton, intended for use as a bandage over the eye for protection or absorption of secretions. It is supplied sterile to allow for use in the operating room, and for application to the eye following injury and o
| Date | Mar 27, 2017 (initiated Mar 3, 2017) |
| Status | Terminated |
| Reason | Product sterility is compromised due to breach of sterile barrier |
| Root cause | Packaging process control |
| Action | Medtronic/Covidien) initiated recall by letter on March 3, 2017 via Federal Express. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Contact Medtronic representative or Customer Service at (800) 882-5878 For questions regarding this recall call 203-492-5000. |
| Quantity | 24765540 |
| Distribution | Worldwide Distribution. US (Nationwide), Internationally to Canada, Australia, Brunei, Japan, Mexico, Philippines, Singapore, Andorra, Belgium, Botswana, Czech Republic, Denmark, Egypt, France, Greece, Hungary, Ireland, Israel, Italy, Kenya, Libya, Luxembourg, Mauritius, Namibia, Netherlands, Norwa |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.