Recall Z-1497-2017 ยท DYB Introducer, Catheter

Merit Medical Systems, Inc.

7F Prelude Short Sheath Introducer. Catalog Numbers: K15-00070, K15-00170, PSS-7F-4-035MT, PSS-7F-4MT. Produc Usage: The Merit Prelude Short Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The device can also provide access

DateMar 24, 2017 (initiated Feb 16, 2017)
StatusTerminated
ReasonMerit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude Short Sheath Introducers because the sheath marker tips may detach during use.
Root causeProcess control
ActionConsignees notified on 2/16/17 to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact your Merit Sales Representative or Merit Customer Service at 801.208.4381.
Quantity1,265 total
DistributionNationwide and other foreign consignees. No Canadian or VA/govt/military consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082063PRELUDE SHORT SHEATH INTRODUCER
Merit Medical Systems, Inc., Oct 15, 2008