Recall Z-1403-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI.

DateMar 3, 2017 (initiated Apr 4, 2016)
StatusTerminated
ReasonWhen taking measurements from images on the Cardio workstation or from the US cart, numbers are not crossing to the report in the correct unit of measure.
Root causeSoftware design
ActionMerge sent an Urgent : Medical Device Recall letter dated March 28, 2016. If no email address was available, the letters were issued via certified mail. Nonresponders were issued a second letter dated July 11, 2016. For further questions, please call (877) 741-5369.
Quantity110 sites potentially have the affected versions
DistributionDistribution was made to medical facilities nationwide. There was no foreign/military/government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051649VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
Camtronics Medical Systems, Aug 2, 2005