Recall Z-0317-2018 ยท FRN Pump, Infusion
Repro-Med Systems, Inc.
Freedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275, F600, F45, F1200 F180.
| Date | Jan 5, 2018 (initiated Mar 10, 2016) |
| Status | Terminated |
| Reason | On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined t |
| Root cause | Nonconforming Material/Component |
| Action | On March 10, 2016 RMS Medical Products distributed Urgent Voluntary Medical Device Corrections and Removal notices to their customers. On March 12, 2016 & on March 18, 2016, RMS Medical Products distributed a 2nd Urgent Voluntary Medical Device Corrections and Removal notices to their customers advising them that affected products list was reduced after a comprehensive risk assessment was performed. Customers are advised to check stock, quarantine, cease distribution and remove product from use. If customers further distributed product, they are also encouraged to inform the affected partie |
| Distribution | Nationwide in US |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K933652 | FREEDOM 60 SYRINGE INFUSION PUMP SYSTEM WITH SYRINGE SET & TUBING Repro-Med Systems, Inc., May 18, 1994 |