Recall Z-1469-2017 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems

DateFeb 27, 2017 (initiated Feb 21, 2017)
StatusTerminated
ReasonOEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
Root causeProcess control
ActionOn February 21, 2017, an Urgent Medical Device Correction letter was sent to one consignee with an affected unit; a field service representative hand delivered the customer communication.
Quantity1
DistributionOne consignee in CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003837FLEXIVIEW 8800 DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 12, 2001