Recall Z-1483-2017 ยท LXH Orthopedic Manual Surgical Instrument

Acumed LLC

.045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diameter. .054 X 7 ST Guide Wire, Part number 45-0031-S, Sterile 10 Pack of WS-1407ST and has a 7 inch length and 0.054 inch diameter. These guide wires are an orthopedic manual surgical instrument.

DateMar 16, 2017 (initiated Feb 17, 2017)
StatusTerminated
ReasonGuide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).
Root causeProcess control
ActionAcumed sent an Urgent Medical Device Recall letter dated 02/16/2017, to all affected customers on 02/17/2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessService
Quantity24 units of .045 X 6 ST Guide Wire and 25 units of .054 X 7 ST Guide Wire
DistributionWorldwide Distribution in Great Britain, Ireland, and Japan only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.