Recall Z-1769-2017 ยท DWE Tubing, Pump, Cardiopulmonary Bypass

Datascope Corporation

Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung

DateMar 23, 2017 (initiated Feb 16, 2017)
StatusTerminated
ReasonA complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
Root causeDevice Design
ActionMaquet Inc. was notified by their customer on January 1, 2017. This was a reverse notification. The one impacted customer said they will not return the product since they can easily correct the pigtail connection before use. For further questions regarding this recall, please call (973) 709-7442.
Quantity21 kits
DistributionUS Distribution to the state of : CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080592HLM TUBING SET WITH BIOLINE COATING
Maquet Cardiopulmonary, AG, Sep 11, 2008