Recall Z-2179-2017 ยท IXR Table, Radiographic, Tilting

GE Healthcare, LLC

Table Patient Step on RFX/SFX, Legacy, and Precision 500D systems

DateMay 22, 2017 (initiated Feb 6, 2017)
StatusTerminated
ReasonReported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator.
Root causeDevice Design
ActionConsignees were sent on 2/6/2017 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10908 dated February 3rd, 2017. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator & Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact consignee to arrange for the correction.. For questions contact GE Healthcare Service at 1-80
Quantity2,380 (2,215 US; 165 OUS)
DistributionUS: Nationwide including DC and Puerto Rico. OUS: Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt, France, Germany, Greece, Honduras, Hungary, Indonesia, Ireland, Israel, Italy, Jamaica, Jordan, Kuwait, Libya, Malaysia, Malta, Mexico, Panama, Portugal, Russia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001
K081091PRECISION 500D R&F X-RAY SYSTEM
GE Healthcare, Jun 11, 2008
K760466TABLE, X-RAY, RFX CLASSICAL
General Electric Co., Aug 26, 1976
K801320SFX-90 TABLE/SPOTFILMER W/FLUORICON L200
General Electric Co., Jul 21, 1980
K862120SCX SYSTEM
General Electric Co., Jun 16, 1986
K973039LEGACY/LEGACY-D TABLE
GE Medical Systems, Oct 16, 1997