Recall Z-1858-2017 ยท JDS Nail, Fixation, Bone

Zimmer Biomet, Inc.

various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant.

DateMar 27, 2017 (initiated Feb 17, 2017)
StatusTerminated
ReasonEndotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
Root causeProcess design
ActionOn February 17, 2017 URGENT MEDICAL DEVICE RECALL (REMOVAL) - LOT SPECIFIC notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Distributors, Sales Representatives, and Distributor Operation Managers Your Responsibilities: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. C
Quantity28,253 in total
DistributionWorldwide - US Nationwide in the ARGENTINA AUSTRALIA CANADA CHILE CHINA COSTA RICA ECUADOR GSCC INDIA ITALY JAPAN KOREA, THE REP OF (S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K800144FEMORAL CEMENT RESTRICTOR
Biomet, Inc., Feb 19, 1980