Recall Z-1414-2017 ยท JQP Calculator/Data Processing Module, For Clinical Use

Becton Dickinson & Co.

EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting feature

DateFeb 27, 2017 (initiated Aug 25, 2014)
StatusTerminated
ReasonThe information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.
Root causeVendor change control
ActionCustomers were initially contacted via phone calls and subsequently provided with a written communication. For questions contact BD Technical Services Department at 1-800-638-8663.
QuantityQty Distr. to Field - 115
Distribution15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
FDA recordOpen on fda.gov

510(k) clearances named in this recall