Recall Z-1394-2017 ยท MLA Dna Probe, Yeast

Becton Dickinson & Co.

BD AffirM VPIII Microbial Identification Test

DateFeb 27, 2017 (initiated Aug 27, 2015)
StatusTerminated
ReasonBD has confirmed that a portion of tests associated with the affected lots( 446252 and 446257 ) of product may exhibit blue speckles or fibers on the beads and cards, as shown below. This issue would result in an invalid test, as described in the package insert.
Root causeEmployee error
ActionBecton Dickinson sent an Urgent Product Recall Letter dated August 2015. The distributors were provided with a written communication via email and customers were provided with a written communication via UPS. Customers were not asked to send product back. If further assistance is needed regarding replacements, please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department at 1-800-638-8663. For further questions please call (410) 316-4428.
QuantityQty Distributed to Field - 4,502
DistributionWorldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY and t
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K931374AFFIRM VPIII MICROBIAL IDENTIFICATION TEST
Microprobe Corp., Jun 15, 1993