ClearanceBrief
Recall Z-1413-2017 ยท HWA Impactor

Encore Medical, Lp

RSP Impaction Fixture

DateFeb 9, 2017
StatusTerminated
ReasonDuring the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Root causeDevice Design
ActionThere are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.
Quantity626 units
DistributionUS, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.