ClearanceBrief
Recall Z-1433-2017 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

MEVATRON M 6700 (MX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer

DateJan 25, 2017
StatusTerminated
ReasonSiemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
Root causeUnder Investigation by firm
ActionSiemens sent a Customer Safety Advisory Notice (CSAN) dated February 7, 2017, to affected customers to inform them of a planned hardware update regarding the mechanics of the overhead suspension of their Digital Linear Accelerator. The notice also explained the actions Siemens is taking to perform the necessary fix at the affected sites. Customers were advised to notify anyone at their organization that should be aware of this information. Also, the safety notice should be forwarded to the new owner and Siemens also requested they be informed of the identity of the device's new owner where
Quantity35 systems
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790688MEVATRON 67
Siemens Medical Laboratories, Inc., Apr 23, 1979