Recall Z-1212-2017 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ARTISTE MV System, medical charged-particle radiation therapy system Product Usage: The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer.

DateJan 25, 2017
StatusTerminated
ReasonSiemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
Root causeDevice Design
ActionSiemens mailed a Customer Safety Advisory Notice (CSAN) dated February 7, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that Siemens will contact customers to arrange a date for the installation of the software update.
Quantity194 systems
DistributionUnited States distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072485ARTISTE MV SA
Siemens Medical Solutions USA, Inc., Dec 27, 2007