Recall Z-1464-2017 ยท FPA Set, Administration, Intravascular

CareFusion 303, Inc.

MaxGuard Extension Set with Injection Site and 0.2 micron filter, Model No. ME3305

DateFeb 28, 2017 (initiated Feb 20, 2017)
StatusTerminated
ReasonCareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing.
Root causeProcess control
ActionAn Urgent Medical Device Recall Notification letter will be sent to customers on February 20, 2017 to inform them that CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter model number ME3305 Lot Number 16016790. Customers are informed that CareFusion has received reports of separation and/or leakages between theY-connector and tubing. Customers are informed that leakages from an extension set can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, or underinfusion. The letter informs the customers of the
Quantity1,200 units
DistributionUS Distribution including MA and MD
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051499INTRAVASCULAR ADMINISTRATION SET AND EXTENSION SET
Medegen Medical Manufacturing Services, Jun 22, 2005