ClearanceBrief
Recall Z-1235-2017 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Terumo Cardiovascular Systems Corporation

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.

DateFeb 22, 2017 (initiated May 27, 2015)
StatusTerminated
ReasonTerumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is
Root causeProcess control
ActionCustomer Notification Letter/Response Form is being sent to all affected consignees.
Quantity579 units have entered into the distribution chain
DistributionTerumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112229SARN CENTRIFUGAL PUMP
Terumo Cardiovascular Systems, Dec 9, 2011