ClearanceBrief
Recall Z-1578-2017 ยท LRO General Surgery Tray

Resource Optimization & Innovation Llc

Regard, Item Number: 880342004, Sterile, GS0776D - Laparotomy - OSU

DateJan 9, 2017
StatusTerminated
ReasonCustom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.
Root causeNonconforming Material/Component
ActionROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793.
Quantity30 kits
DistributionUS Distribution to LA and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall