Recall Z-1187-2021 ยท JZO System, Test, Thyroid Autoantibody

Tosoh Bioscience Inc

Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that t

DateFeb 23, 2012
StatusTerminated
ReasonSome combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.
Root causeNonconforming Material/Component
ActionOn 7/24/2012 the firm sent recall notificaiton letters to their customers.
Quantity34 boxess/2 sets/6 bottles
DistributionUS Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.