Recall Z-1428-2017 ยท KXJ Table, Radiologic

Villa Sistemi Medicali S.P.A.

Apollo; Models: 9784000131 9784000231 9784020131 9784020231 9784100231 9784120131 9784120231 9784130131 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.

DateDec 27, 2016
StatusTerminated
ReasonPatient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Root causeNonconforming Material/Component
ActionCustomers were notified of the recall via DHL or similar mail service. The letter identified the affected product, problem and actions to be taken.
Quantity65 devices in total
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050190APOLLO
Villa Sistemi Medicali S.P.A., Mar 25, 2005