Recall Z-1693-2017 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2

DateJan 26, 2017
StatusTerminated
ReasonSIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.
Root causeProcess design
ActionURGENT MEDICAL DEVICE CORRECTION Letters and Reply Forms (dated 1/26/2017) were mailed to customers via USPS first-class mail on 1/26/2017 providing the necessary instructions to customers. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
QuantityDomestic: 14,175 units; Canada: 187 units
DistributionNatiowide, Puerto Rico and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133801BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Baxter Healthcare Corporation, May 7, 2014