Recall Z-1382-2017 ยท KKY Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction

Matrix Surgical Holdings, LLC

OP7543 Design Y Mandible Onlay - Medium - Left and OP7544 Design Y Mandible Onlay - Medium - Right

DateFeb 28, 2017 (initiated Jan 25, 2017)
StatusTerminated
ReasonMislabeling. The packaging labeled as left mandible contained a right mandible.
Root causeOther
ActionMatrix Surgical sent an Urgent Product Recall e-mail dated January 26, 2017, to the foreign consignees. The distributors were instructed to return the product in inventory, and to conduct a sub-recall, if needed. For further questions, please call ( 404) 862-3794.
Quantity28
DistributionWorldwide Distribution to the countries of : Brazil, Chile, Colombia, Norway, Saudi Arabia, South Africa, Sweden and Taiwan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123908OMNIPORE SURGICAL IMPLANTS AND ACCESSORIES
Matrix Surgical USA, Jul 31, 2013
K133046OMNIPORE CUSTOMIZED SURGICAL IMPLANTS
Matrix Surgical Holdings, LLC/Matrix Surgical USA, May 14, 2014