Recall Z-1201-2017 ยท MWS Stabilizer, Heart

Terumo Cardiovascular Systems Corporation

Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel

DateFeb 14, 2017 (initiated Aug 20, 2014)
StatusTerminated
ReasonTerumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.
Root causeDevice Design
ActionTerumo Cardiovascular Systems has contact all affected parties (user facilities and consignees) via a Customer Notification Letter that advises of the recall activity.
QuantityThe total number of distributed devices stands at 260
DistributionTerumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.