Recall Z-1510-2017 ยท FMZ Incubator, Neonatal

Ohmeda Medical

CarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for neonates who are unable to provide their own thermoregulation. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. Most Incubators provide a controlled thermal environment for neonates who are unable to provide th

DateNov 18, 2016
StatusTerminated
ReasonCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Root causeUnder Investigation by firm
ActionAccording to the firm, this issue will be resolved by sending an Urgent Medical Device Correction customer letter to customers with affected Power Cords. This letter will inform the customer of the issue and provide safety instructions for immediate mitigation of the issue. The long-term solution is to re-place all affected Power Cords at no cost to the customer. According to the firm, the plan for public notification is to contact all customers with affected Power Cords by means of an Urgent Medical Device Correction letter. In the US this will be sent by a traceable means (either US Postal
Quantity5,993 (5,080 US; 913 OUS)
DistributionWorldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974349OHMEDA-OHIO CARE PLUS INCUBATOR
Ohmeda Medical, Feb 12, 1998