Recall Z-1501-2017 ยท KZF Device, Medical Examination, Ac Powered
Ohmeda Medical
Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical fi
| Date | Nov 18, 2016 |
| Status | Terminated |
| Reason | Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury. |
| Root cause | Under Investigation by firm |
| Action | According to the firm, this issue will be resolved by sending an Urgent Medical Device Correction customer letter to customers with affected Power Cords. This letter will inform the customer of the issue and provide safety instructions for immediate mitigation of the issue. The long-term solution is to re-place all affected Power Cords at no cost to the customer. According to the firm, the plan for public notification is to contact all customers with affected Power Cords by means of an Urgent Medical Device Correction letter. In the US this will be sent by a traceable means (either US Postal |
| Quantity | 5,993 (5,080 US; 913 OUS) |
| Distribution | Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.