Recall Z-1180-2017 ยท LLZ System, Image Processing, Radiological
Merge Healthcare, Inc.
Merge RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.
| Date | Feb 9, 2017 (initiated Nov 17, 2015) |
| Status | Terminated |
| Reason | The values provided from the Pixel Value tool do not appear to be correct, which may result in potential patient injury or delay in diagnosis or treatment. |
| Root cause | Software design |
| Action | Merge sent an Urgent Medical Device Recall letter dated March 28, 2016 to all affected customers. The letter notifies the customer of the issue and informs them an upgrade is available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provid |
| Quantity | 8 sites have the potentially affected version |
| Distribution | US Distribution was made to medical facilities in AL, IN, MI, PA and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K053281 | EVMS (ENTERPRISE VISUAL MEDICAL SYSTEM SOFTWARE) Emageon, Inc., Dec 7, 2005 |