Recall Z-1198-2017 ยท HST Apparatus, Traction, Non-Powered

Allen Medical Systems, Inc.

Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shoulder joint during arthroscopic surgical procedures.

DateFeb 14, 2017 (initiated Jan 3, 2017)
StatusTerminated
ReasonDuring use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
Root causeDevice Design
ActionThe distributor was notified of the recall via email on January 3, 2017. The letter identified the affected product, problem and actions to be taken. The letter instructs the consignee to remove the product from service and follow instructions to return the device to Allen Medical for repair.
Quantity112 devices
DistributionUS state of FL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.