Recall Z-1052-2017 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Carestream Health Inc

Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System Diagnostic ultrasound imaging or fluid flow analysis of the human body

DateJan 19, 2017 (initiated Nov 21, 2016)
StatusTerminated
ReasonSoftware error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters. As such, the user expects that the measurement is taken calculated in centimeters, consistent with other values in the report. In actuality, the measurement is taken in millimeters. When this updated measurement is used in an average calculation, the result appears incorrect as two measurements are interpreted as centimeters while the user cor
Root causeSoftware design
ActionCarestream sent an URGENT: MEDICAL DEVICE CORRECTION Letter dated November 17, 2016, with a Consignee Notification Acknowledgement (HRA # MA-2016-019) to all affected US consignees via UPS on November 21, 2016. The letter identified the product, the problem, and the action to be taken by the consignee Action to be Taken: When reviewing the worksheet, if a zero value is observed and that calculation field needs to be changed, the calculation needs to be deleted from the worksheet. Once deleted, the user needs to acquire a new calculation from a clinical image. Carestream is currently validatin
QuantityUS: 14 units; Foreign: 16 units
DistributionWorldwide Distribution - US including SC, GA, TX, OH, IA, NY and foreign distribution to Switzerland and Malaysia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152467Touch Ultrasound
Carestream Health, Inc., Dec 9, 2015