ClearanceBrief
Recall Z-1420-2017 ยท JLS Radioimmunoassay, Progesterone

Siemens Healthcare Diagnostics, Inc

IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in vitro diagnostic use with the IMMULITE2000 Systems Analyzers- for the quantitative measurement of progesterone in serum, as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.

DateJan 4, 2017
StatusTerminated
ReasonDHEA-S causes falsely elevated progesterone results.
Root causeNonconforming Material/Component
ActionSiemens sent an Urgent Medical Device Correction (UMDC) letter dated January 4, 2017 to all affected Siemens Healthcare Diagnostics customers in the United States dated . An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 4, 2017. These letters identified the affected product problem and actions to be taken. For questions contact your local Siemens Customer Care Center or your local Siemens Technical support.
Quantity9076
DistributionUS Nationwide distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932955CIBA CORNING ACS PROGESTERONE IMMUNOASSAY
Ciba Corning Diagnostics Corp., Aug 2, 1993