Recall Z-1419-2017 ยท JLS Radioimmunoassay, Progesterone

Siemens Healthcare Diagnostics, Inc

Dimension Vista LOCI Progesterone; Catalog Number: K6464; Siemens Material Number: 10461743 Product Usage: For in vitro diagnostic use in the quantitative determination of progesterone in serum and using the ADVIA Centaur, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems

DateJan 4, 2017
StatusTerminated
ReasonDHEA-S causes falsely elevated progesterone results.
Root causeNonconforming Material/Component
ActionSiemens sent an Urgent Medical Device Correction (UMDC) letter dated January 4, 2017 to all affected Siemens Healthcare Diagnostics customers in the United States dated . An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 4, 2017. These letters identified the affected product problem and actions to be taken. For questions contact your local Siemens Customer Care Center or your local Siemens Technical support.
Quantity1148
DistributionUS Nationwide distribution including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932955CIBA CORNING ACS PROGESTERONE IMMUNOASSAY
Ciba Corning Diagnostics Corp., Aug 2, 1993