Recall Z-1418-2017 ยท JLS Radioimmunoassay, Progesterone
Siemens Healthcare Diagnostics, Inc
ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro diagnostic test for the quantitative measurement of progesterone in human serum and plasma on the Dimension VistaSystem. Progesterone measurements are used in the diagnosis and treatment of disorders of the ovaries or placenta.
| Date | Jan 4, 2017 |
| Status | Terminated |
| Reason | DHEA-S causes falsely elevated progesterone results. |
| Root cause | Nonconforming Material/Component |
| Action | Siemens sent an Urgent Medical Device Correction (UMDC) letter dated January 4, 2017 to all affected Siemens Healthcare Diagnostics customers in the United States dated . An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 4, 2017. These letters identified the affected product problem and actions to be taken. For questions contact your local Siemens Customer Care Center or your local Siemens Technical support. |
| Quantity | 76,107 |
| Distribution | US Nationwide distribution including Puerto Rico |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K932955 | CIBA CORNING ACS PROGESTERONE IMMUNOASSAY Ciba Corning Diagnostics Corp., Aug 2, 1993 |