Recall Z-1197-2017 ยท MLR Sterilizer, Chemical

Advanced Sterilization Products

STERRAD NX Sterilization System, Product Code: 10033, 10033-002

DateFeb 14, 2017 (initiated Jan 9, 2017)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.
Root causeSoftware design
ActionAdvanced Sterilization will send an Field Safety Notice dated January 9, 2017,to inform customers that Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration. ASP is sending the letter to remind customers to follow the instructions in the STERRAD NX System Users Guide. Specifically, after cycle initiation, the user should only open the chamber door of the STERRAD NX System when prompted by the Graphical User Interface (GUI) pe
Quantity7,694 units total (5,935 units in US)
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Australia, China, Hong Kong, India, Indonesia, Korea (South), Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Taiwan, Thailand, Vietnam, Canada, Belgium, Czech Republic, Egypt, France, Germany, Great
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042116STERRAD NX STERILIZER
Advanced Sterilization Products, Apr 22, 2005
K142454STERRAD NX Sterilizer, STERRAD 100NX Sterilizer
Advanced Sterilization Products, Apr 3, 2015
K151725STERRAD NX Sterilizer, STERRAD 100NX Sterilizer
Advanced Sterilization Products, Jan 13, 2016