ClearanceBrief
Recall Z-1163-2017 ยท NRA Prosthesis, Knee, Femorotibial, Unicompartmental, Semi-Constrained, Metal/Polymer, Mobile Bearing

Biomet U.K., Ltd.

Oxford Partial Knee System Left Medial Tibial Trial Tray Size D Product Usage: Instruments for use with the Oxford Uni Partial Knee System

DateJan 5, 2016
StatusTerminated
ReasonZimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
Root causeNonconforming Material/Component
ActionZimmer Biomet initiated a voluntary recall of the Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the ffected products by mailing letters via FedEx on 01/05/2017. Customers were instructed to Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in their inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form. a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within thr
Quantity72 units in total
DistributionWorldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.