Recall Z-0989-2017 ยท NEP System/Device, Pharmacy Compounding

HEALTH ROBOTICS S.R.L.

i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448

DateJan 10, 2017 (initiated Nov 23, 2016)
StatusTerminated
ReasonThe firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.
Root causeSoftware design
ActionOmincell mailed an Urgent Medical Device Field Correction Notification letter to affected customers on 11/21/2016. The letter provided information of the event and instructions on how to continue operating the device to prevent risks. The issue will be corrected with a software update.
Quantity121 systems
DistributionShipped throughout United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall