Recall Z-1969-2017 ยท BWN Table And Attachments, Operating-Room

The Anspach Effort, Inc.

9.4CM MICRO REVISION ATTACH, BLACK MAX

DateOct 6, 2016
StatusTerminated
ReasonSupplied Directions for Use (DFU) did not contain a recommended service interval.
Root causeUnder Investigation by firm
ActionConsignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees
Quantity11
DistributionWorldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955084ASPACH 65,000 (MODIFICATION)